Table of Contents

Tube oven annealing

The tube oven has a temperature range of 50-1200 °C and can fit 1 inch diameter tubes. Temperature regulation is done by a Eurotherm 2404 controller. Three different process gasses can be connected, at the moment H2 and Ar are connected. Gas flow of each gas is controlled by a mass flow controller (MFC), mixing is possible. A dry vacuumpump is used to pump down the system, lowest pressure that can be achieved is roughly 1e-2mbar. A pressure control valve can be used to control either the flow or pressure in the tube. Most often the pressure control valve is operated in the fully open state to achieve a pressure as low as possible. Pressure can be controlled between 1e-2mbar and 100mbar.

A Python program on the PC “hydrogen_annealing” can be used to set gas flows, the pressure set point and monitor and log the process. To ensure safe handling of hydrogen, mulitple safety interlocks are present. If an interlock condition is not met, the main supply at the H2 gas cylinder will be closed automatically.

Temperature regulation of the oven is done with Eurotherm software; “iTools engineering studio”.

Notes before usage

Tube specifications

Outer diameter: 25mm or 1 inch (25.4mm)
Length: between 62cm and 70cm

Operating temperatures:

Safety

Most important safety points are:

Hydrogen safety

If a process requires hydrogen we need to make sure it doesn't escape into the lab and the gas is mixed with nitrogen at the inlet of the pump to achieve a H2/N2 mixture below LEL.

Flammable range of H2 in air is >4% (LEL) and <75%(UEL).

A flammable mixture of H2 and Air can be ignited on a hot surface if the temperature exceeds 500°C (in literature different values are mentioned, 500°C is chosen on the safe side).
Hence when working with Hydrogen do not use processes above this temperature.

There should be adequate ventilation in the lab, the window should be open.

no sources of ignition close to the setup

Dangerous situations can occur when:

H2 safety interlocks

There is a safety valve directly behind the hydrogen cylinder in the safety cabinet, if one of the interlocks is not met the valve will close automatically.
When all safety interlocks are met, the valve can be openend manually from the GUI.

  1. h2 level in lab < 2%
  2. n2 flow > 100*h2 flow and 100*h2 setpoint
  3. fume hood exhaust = on
  4. vacuum pump in normal operation
  5. leak test interlock: first achieve base pressure
    • if flow is on: pressure should stay below 100mbar
    • when interlock has triggered, the interlock can be reset after achieving base pressure
  6. connection with PLC

High temperature safety

Always lock the oven when a process is running. Let process cool down to 50°C before handling any components. No organic solvents on the table when a process is running. Work with clean tubes

Pressurized gas cylinders safety

Do not attempt to change cylinders or connected equipment, always consult with a technician.

Setting up and starting the process

Cool down and finish

Do the following steps ASAP if your process allows for it while the oven cools down, if you need gas flow when cooling down the sample; first turn of the oven and wait until the sample has cooled down to the desired temperature before following the steps below

Common problems

PLC not connected

The extra network card in the PC sometimes loses connection. Open the PC and unplug and plug back the network card. Restart the computer. Open the PLC program and try to connect to the PLC. LOGO!SoftComfort will recognize that the IPadres of the PLC is not in the same subnet mask and will ask to add this; acknowledge. Now you should have connection again.