===== Plasma Cleaner manual ==== {{ :plasma_cleaner_pdc-32g-2_usermanual_2022.pdf |Plasma Cleaner manual}} ==== DO's and DONT's ==== Only use bare substrate in the plasma cleaner, all exceptions should be discussed. * Si * SiO2 * glass/quartz ==== MANUAL ==== * Load samples on the sample tray * Start the vacuum pump while holding the door, make sure black valve is pointing down * Pump down to <100mTorr * Set black valve to the needle valve * if needle valve is open the pressure should go to previous setpoint (check the logbook) * or change the setpoint * Wait for pressure to stabilize * set RF (low, mid or high) * When plasma ignites start your timer * Turn off RF when timer finishes * Close black valve (valve pointing down) * Pump down to <100mTorr to make sure etch products are removed from the chamber * Stop the vacuum pump * Turn black valve to the right to vent * vent slowly to avoid samples flying around in the chamber * Close black valve (valve pointing down) * close the door * write in the logbook: name, date, sample, process conditions, any remarks ==== Recipes ==== Typical cleaning process:\\ 500-900mTorr\\ 1-3 minutes\\ High RF === Maintenance === Tip seal change of vacuum pump\\ Cleaning sample tray and chamber\\