====== MPMS-5S Helium transfer ====== The Squid should be refilled if Helium levels are below 50% ===== Filling ===== - Make sure you have a helium vessel that still has enough (what is enough?) helium - Close all valves of the vessel - Attach hose to compressor and vessel. -{{images:squid-2-compressor.png}} - Open valves between vessel and compressor - Remove fillcap from SQUID. It will POP out! - Attach the fill hose (recovery line) between the SQUID and Cryostat. -{{images:squid-2-valves.png}} - Open the valves on this hose. - Get the large **main** fill line -{{images:squid-2-helium.png}} - Open red valve on He-vessel. Put the large part of the hose into the vessel at average speed. - Wait until cold gas sputters out the other end of the fill hose (~one minute). You will get a cold flame. - Remove the small cap from the recovery line on the SQUID - Insert the other end of the fill hose in the SQUID. (use heat-gun if it gets suck because it got frozen. - Go to the computer and select: utilities->helium fill->start - Plug in the power line of the compressor. - You are now pumping. It is best to monitor the refill. (It will take about 30 minutes) Pay attention to: - The pressure (read-out on He-Vessel indicator) should not exceed 0.5 Bar. You can regulate this by clamping the compressor bypass. - The fill rate (read-out on computer) should be around 1-1.5%/min. - The Vessel Valve should be heated while filling with the heating gun(at half-heating power), to prevent the valve from freezing. - There are no instructions yet to stop the filling and disconnect the He-Vessel. Ask for help. Original (InkScape) {{wiki:squid.svg.pdf|vector graphic}} (remove pdf extension).